Contact an Account ManagerDetermination of bromhexine in plasma by reversed phase liquid chromatography. Interference of lipoproteins on extraction
Summary
Type: HPLC
Application #: 105120
Catalog#: 720120.46 - MN HPLC Column, Nucleosil C18, 100A, 5um, 4.6x150mm
Detection: UV, 308nm
Conditions: eluent: 5mM dimethylhexylamine in methanol - phosphate buffer pH 2 (59:41, v/v
flow rate 1.0ml/min
Authors: Johansson, M. et al.
Source: J. Chromatogr. 432 (1988) 243–252
Details
Legend: Fig. 5. Comparisons of plasma chromatograms containing 34ng/ml bromhexine A)
without and B) with injection of a step gradient of 500µl of 5mM
dimethylhexylamine in 90% (v/v) methanol in phosphate bu